宇航计测技术 ›› 2023, Vol. 43 ›› Issue (5): 68-74.doi: 10.12060/j.issn.1000-7202.2023.05.11

• 量值传递技术 • 上一篇    下一篇

环境核辐射监测仪表剂量测量比对

高飞,王靖谦,董璐琪,王菲菲   

  1. 中国原子能科学研究院,中核核工业计量与测试技术重点实验室,计量与校准技术重点实验室,北京 102413
  • 出版日期:2023-11-15 发布日期:2023-12-09
  • 作者简介:高飞(1983-),男,研究员,博士,主要研究方向:辐射防护及电离辐射计量技术。

Intercomparison of Environmental Nuclear Radiation Measuring

GAO Fei,WANG Jing-qian,DONG Lu-qi,WANG Fei-fei   

  1. China Institute of Atomic Energy,China National Nuclear Corporation Key Laboratory of Metrology and Measurement Technology,Key Laboratory for Metrology and Calibration Techniques,Beijing 102413,China
  • Online:2023-11-15 Published:2023-12-09

摘要: 为检验和促进提高校准能力及操作人员技术能力水平,确保监测数据的可靠性,开展了环境核辐射监测仪表的测量比对工作。本次比对主要针对环境级剂量率仪、探测下限覆盖环境范围的防护级剂量率仪和γ射线空气比释动能次级标准剂量仪。比对参数涉及空气比释动能和周围剂量当量。此次比对覆盖了广泛配备的环境核辐射监测仪表,对此类仪表的X射线响应、γ射线响应和宇宙射线响应等技术性能进行测量比对,分别采用En值及相对误差对比对结果进行分析评价。由比对结果可知,目前校准实验室所使用的次级标准剂量仪的|En|<1,全部符合要求,大部分环境核辐射监测仪表满足S-Cs各点指标要求,但是X射线响应较差,部分监测仪表不能满足能量响应变化极限不超过±30%的要求,建议仪表使用单位在首检时对其能量响应进行检定。

关键词: 环境核辐射, γ射线, 空气比释动能, 比对

Abstract: In order to check and promote the improvement of calibration ability and the technical ability of operators,and ensure the reliability of monitoring data,the measurement and comparison work of environmental nuclear radiation monitoring instrument was carried out.The intercomparison includes environmental level dosemeters, protection level dosemeters which coverage for environmental level,and γ-ray air kerma secondary standard ionization chamber.The intercomparison parameters involve air kerma and ambient dose equivalent.The intercomparison covers a wide range of environmental nuclear radiation monitoring instruments.X-ray response,γ-rays and cosmic rays response properties of the environmental nuclear radiation monitoring instruments were studied.The En value and relative error were compared to analyze and evaluate the results.As the intercomparison results,the En values of standard ionization chambers in calibration laboratory are less than 1,and all meet requirements.Most of the environmental radiation monitoring instruments meet the requirements of S-Cs measurement.However,the X-ray response of most monitoring instruments are poor.Some monitoring instrument cannot meet the requirements that the energy response variation limit dose not exceed ±30%,so it is recommended that the instrument user should verify its energy response during the first inspection.

Key words: Environmental nuclear radiation, γ-rays, Air kerma, Intercomparison