宇航计测技术

• 论文 • 上一篇    下一篇

近红外绝对光谱响应度传递标准的研制

刘玉龙1;刘长明1;史学舜1;陈振林1;李健军2   

  1. 1、中国电子科技集团公司第41研究所,山东青岛 266555;
    2、中科院安徽光学精密机械研究所,安徽合肥 230031
  • 出版日期:2015-06-15 发布日期:2015-06-15
  • 作者简介:刘玉龙(1988-), 男,工程师,硕士,主要研究方向:光电探测器参数计量技术。

Development of the Transfer Standard of Near Infrared Absolute Spectral Response

LIU Yu-long1;LIU Chang-ming1;SHI Xue-shun1;CHEN Zhen-lin1;LI Jian-jun2   

  1. 1、The 41st Institute of China Electronics Technology Group Corporation, Qingdao, Shandong 266555;
    2、Anhui institute of optical precision machinery, Chinese academy of sciences, Hefei, Anhui 230031
  • Online:2015-06-15 Published:2015-06-15

摘要: 在光辐射计量中,陷阱探测技术是一种重要的探测技术,陷阱探测器是光辐射功率基准量值保持和传递的主要媒介。InGaAs陷阱探测器常被作为(900~1 700)nm波段光辐射计量的传递标准。本文基于陷阱探测技术,针对近红外波段量值传递的需求,研制了InGaAs陷阱探测器,并进行了性能测试。测试结果显示:在1 064nm处绝对光谱响应率为0.709 8A/W,测量不确定度达到0.1%,在(0.1~3)mW内,其非线性指标优于0.999 93,1h测试非稳定性为0.004 95%,空间非均匀性达到0.05%,满足作为传递标准的技术要求。

关键词: 近红外, 绝对光谱响应度, 传递标准, +InGaAs陷阱探测器

Abstract: In optical radiation measurement, trap detection technology is an important detecting technology,and the trap detector is a major methods of keeping and transferring optical radiation power benchmark value. InGaAs trap detector is often as optical radiation transfer standard in 900nm to 1 700nm. In view of the requirements of dissemination in the near infrared band, InGaAs trap detector is developed by the trap detection technology, and the performance test was performed in this article. Test results show: the absolute spectral response rate of the trap detector was 0.709 8A/W at 1 064nm, and the uncertainty of measurement is 0.1%; The nonlinear index of the spectral response is better than 0.999 93 within the 0.1mW~3mW; The non stability of the measurement results is better than 0.004 95% in an hour; The spatial heterogeneity of the trap detector is better than 0.05%, All performance parameters from the tests meet the technical requirements as a transfer standard.

Key words: Near-infrared, Absolute spectral response, Transfer standard, +InGaAs trap detector