JOURNAL OF ASTRONAUTIC METROLOGY AND MEASUREMENT ›› 2015, Vol. 35 ›› Issue (1): 1-4.doi: 10.12060/j.issn.1000-7202.2015.01.01

    Next Articles

Experiments Research of Calibration of Sampling Oscilloscope Risetime based on Optic-electrical Pulse

GONG Peng-wei1;JIANG He1;CHEN Bei1;CHEN Dong-qing1;XIE Wen1;MA Hong-mei11;YANG Chun-tao1,2   

  1. 1、Beijing Institute of Radio Metrology and Measurement, Beijing 100039;
    2、Key Laboratory of Science and Technology for National Defence on Metrology and Calibration Technology,Beijing 100039
  • Online:2015-02-15 Published:2015-02-15

Abstract: The paper introduces calibration method of sampling oscilloscope risetime based on optic-electrical pulse and constructs calibration setup. The risetime of sampling oscilloscope with 30GHz bandwidth and 50GHz bandwidth is calibrated by using this calibration setup. According to experiment results, optic-electrical pulse is suitable for the calibration of sampling oscilloscope risetime with Gaussian response. According to this method, Beijing Institute of Radio Metrology and Measurement constructed calibration setup.

Key words: Calibration, Risetime, Optic-electrical pulse, Oscilloscope