JOURNAL OF ASTRONAUTIC METROLOGY AND MEASUREMENT ›› 2012, Vol. 32 ›› Issue (5): 41-44.doi: 10.12060/j.issn.1000-7202.2012.05.09
Previous Articles Next Articles
WU Kang1;LI Ya-lu1;LIU Min1
Online:
Published:
Abstract: The article analysis and calculate uncertainty of 10μF~1mF high capacitance’s measurement results by series capacitor method which standard are high accuracy bridge AH2700A and standard capacitor Agilent 16380C(1??F).Via uncertainty analysis, measurement uncertainty of capacitor of 10??F is 0.02%,measurement uncertainty of capacitor of 100??F is 0.1%, the measurement uncertainty of capacitor of 1mF is 0.6%.
Key words: Series capacitor method, High capacitance standard, Uncertainty analysis
WU Kang;LI Ya-lu;LIU Min. Uncertainty Analysis of High Capacitance Calibration by Series Capacitor Method[J]. JOURNAL OF ASTRONAUTIC METROLOGY AND MEASUREMENT, 2012, 32(5): 41-44.
0 / / Recommend
Add to citation manager EndNote|Ris|BibTeX
URL: http://www.yhjcjs.com.cn/EN/10.12060/j.issn.1000-7202.2012.05.09
http://www.yhjcjs.com.cn/EN/Y2012/V32/I5/41