Journal of Astronautic Metrology and Measurement ›› 2024, Vol. 44 ›› Issue (4): 35-39.doi: 10.12060/j.issn.1000-7202.2024.04.06

Previous Articles     Next Articles

Test and Analysis of Thermal Response Time of Temperature Sensor

LIU Baoju1,GUO Yafei1,ZHANG Junqi1,ZHANG Zhongjie3,LV Yueyuan1,LIU Lin1,CHEN Yiqing1   

  1. 1.Beijing Aerospace Institute for Metrology and Measurement Technology,Beijing 100076,China;
    2.State-owned Luoyang Dancheng wireless Power Plant Luoyang 471000,China;
    3.Beijing Institute of Space Long March Vehicle,Beijing 100076,China
  • Online:2024-08-25 Published:2024-09-14

Abstract: Most of the domestic thermal response time tests for temperature sensors are conducted in 90 ℃ silicone oil by oil bath method,and the test method is single without considering the relationship between the actual use environment of temperature sensors and the test environment,and the test method is manual test with low efficiency and poor reproducibility.To solve this problem,an automatic test device for thermal response time of temperature sensor is designed based on Labview software.The thermal response time of PT100 platinum resistor,PT1000 platinum resistor and thermistor temperature sensor is tested,and the influence of different packaging and heating methods on the thermal response time of temperature sensor is analyzed.It is proved that different packaging methods and heating methods have great influence on the thermal response time test.The test device studied can be applied to the response time measurement of all types of temperature sensors.

Key words: Temperature sensor, Thermal response time, Dynamic characteristics, Automatic test equipment

CLC Number: